Performance study of the automated immunoassay test anti-hepatitis C virus VIDAS® for the qualitative detection of antibodies anti-hepatitis C virus
Materials and methods: Routine anti-HCV screening of clinical samples was carried out on the ARCHITECT® i2000SR platform (Abbot Laboratories, Abbott Park, IL, USA); out of 8876 samples tested from October 2012 to May 2013, 70 sera with low positive anti-HCV results (1≤S/CO<8) were collected. These samples were tested for the presence of anti-HCV antibodies using anti-HCV VIDAS® and INNO-LIATM HCV score assays (Innogenetics NV, Ghent, Belgium).
Results and conclusions: Positive anti-HCV results were obtained in 61.4% and 41.4% of sera tested with VIDAS® and INNO-LIATM, respectively. Concordance between methods was 63.2% for ARCHITECT® and VIDAS®, 42.6% for ARCHITECT® and INNO-LIATM, and 79.4% for VIDAS® and INNO-LIATM. Anti-HCV VIDAS® demonstrated superior specificity compared to the anti-HCV test ARCHITECT®; therefore, this assay has been introduced in our routine analysis as a second level screening test to select samples to be subjected to confirmatory anti- HCV immunoblot.
How to Cite
Salvetti, S., Lavinia, F., & Montenora, I. (2016). Performance study of the automated immunoassay test anti-hepatitis C virus VIDAS® for the qualitative detection of antibodies anti-hepatitis C virus. Microbiologia Medica, 31(1). https://doi.org/10.4081/mm.2016.4808