Reliability analysis of GEM® Premier™ technology: a multicenter study

Submitted: 24 November 2014
Accepted: 16 July 2015
Published: 22 October 2015
Abstract Views: 1796
PDF: 909
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Authors

This paper describes how specific blood gas analyzer characteristics can support the emerging clinical and organizational user’s needs while ensuring patient safety. A oneyear data set from two Italian hospitals was analyzed from 10 different blood gas analyzers. Performance measurements in terms of mean down time (MDT) were calculated to show how technical characteristics declared by the manufacturer compare to the analyzer availability in real clinical settings. Results show a high level of reliability for the analyzed technology, associated with very low MDT of each sensor integrated in the cartridge. Moreover, results show a high level of efficiency in cartridge usage. Such results are consistent with the specification of the GEM® Premier™ maintenance- free technology and are particularly relevant in the point-of-care testing setting.

Dimensions

Altmetric

PlumX Metrics

Downloads

Download data is not yet available.

Citations

How to Cite

Faggiano, F., Franchin, T., Ritrovato, M., & Derrico, P. (2015). Reliability analysis of GEM® Premier™ technology: a multicenter study. Emergency Care Journal, 11(2). https://doi.org/10.4081/ecj.2015.4795