Comparative evaluation of RAD 120 in the TORCH diagnostic

Submitted: 17 February 2014
Accepted: 17 February 2014
Published: 31 December 2008
Abstract Views: 734
PDF: 891
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Authors

This work is a comparative clinical trial about the performance of a new automatic system RAD 120, with final fluorescent determination, and ALISEI, a referential system that works in classical ELISA.This comparison show that RAD120 has some hardware and software problems, that determinate false positive dosages, especially in IgM of Toxoplasmosis and Cytomegalovirus. This system is in evaluation too after adjournment, and only after this second test, we could be sure if RAD 120 can be became another referential system in torch diagnosis.

Dimensions

Altmetric

PlumX Metrics

Downloads

Download data is not yet available.

Citations

How to Cite

Potente, G. I., Nisticò, S., Cerminara, M. T., Zangari, E., & Luciano, A. (2008). Comparative evaluation of RAD 120 in the TORCH diagnostic. Microbiologia Medica, 23(4). https://doi.org/10.4081/mm.2008.2563

Similar Articles

You may also start an advanced similarity search for this article.