Comparative evaluation of RAD 120 in the TORCH diagnostic

Authors

This work is a comparative clinical trial about the performance of a new automatic system RAD 120, with final fluorescent determination, and ALISEI, a referential system that works in classical ELISA.This comparison show that RAD120 has some hardware and software problems, that determinate false positive dosages, especially in IgM of Toxoplasmosis and Cytomegalovirus. This system is in evaluation too after adjournment, and only after this second test, we could be sure if RAD 120 can be became another referential system in torch diagnosis.

How to Cite

Potente, G. I., Nisticò, S., Cerminara, M. T., Zangari, E., & Luciano, A. (2008). Comparative evaluation of RAD 120 in the TORCH diagnostic. Microbiologia Medica, 23(4). https://doi.org/10.4081/mm.2008.2563